1.
Int J Pharm ; 648: 123569, 2023 Dec 15.
Article in English | MEDLINE | ID: mdl-37925043

ABSTRACT

A challenge in pressurised metered-dose inhaler (pMDI) formulation design is management of adhesion of the drug to the canister wall, valve and actuator internal components and surfaces. Wall-material interactions differ between transparent vials used for visual inspection and metal canister pMDI systems. This is of particular concern for low greenhouse warming potential (GWP) formulations where propellant chemistry and solubility with many drugs are not well understood. In this study, we demonstrate a novel application of X-ray fluorescence spectroscopy using synchrotron radiation to assay the contents of surrogate solution and suspension pMDI formulations of potassium iodide and barium sulphate in propellants HFA134a, HFA152a and HFO1234ze(E) using aluminium canisters and standard components. Preliminary results indicate that through unit life drug distribution in the canister valve closure region and actuator can vary significantly with new propellants. For solution formulations HFO1234ze(E) propellant shows the greatest increase in local deposition inside the canister valve closure region as compared to HFA134a and HFA152a, with correspondingly reduced actuator deposition. This is likely driven by chemistry changes. For suspension formulations HFA152a shows the greatest differences, due to its low specific gravity. These changes must be taken into consideration in the development of products utilising low-GWP propellants.


Subject(s)
Metered Dose Inhalers , Nebulizers and Vaporizers , Administration, Inhalation , Catheters , Aluminum , Suspensions , Aerosol Propellants/chemistry , Hydrocarbons, Fluorinated/chemistry
2.
3D Print Med ; 8(1): 2, 2022 Jan 05.
Article in English | MEDLINE | ID: mdl-34985624

ABSTRACT

BACKGROUND: The global pandemic of novel coronavirus (SARS-CoV-2) has led to global shortages of ventilators and accessories. One solution to this problem is to split ventilators between multiple patients, which poses the difficulty of treating two patients with dissimilar ventilation needs. A proposed solution to this problem is the use of 3D-printed flow splitters and restrictors. There is little data available on the reliability of such devices and how the use of different 3D printing methods might affect their performance. METHODS: We performed flow resistance measurements on 30 different 3D-printed restrictor designs produced using a range of fused deposition modelling and stereolithography printers and materials, from consumer grade printers using polylactic acid filament to professional printers using surgical resin. We compared their performance to novel computational fluid dynamics models driven by empirical ventilator flow rate data. This indicates the ideal performance of a part that matches the computer model. RESULTS: The 3D-printed restrictors varied considerably between printers and materials to a sufficient degree that would make them unsafe for clinical use without individual testing. This occurs because the interior surface of the restrictor is rough and has a reduced nominal average diameter when compared to the computer model. However, we have also shown that with careful calibration it is possible to tune the end-inspiratory (tidal) volume by titrating the inspiratory time on the ventilator. CONCLUSIONS: Computer simulations of differential multi patient ventilation indicate that the use of 3D-printed flow splitters is viable. However, in situ testing indicates that using 3D printers to produce flow restricting orifices is not recommended, as the flow resistance can deviate significantly from expected values depending on the type of printer used. TRIAL REGISTRATION: Not applicable.

3.
Int J Pharm ; 597: 120341, 2021 Mar 15.
Article in English | MEDLINE | ID: mdl-33545289

ABSTRACT

In this paper we demonstrate that the use of multiple orifices can improve the fine particle fraction (FPF) of pressurised metered-dose inhaler solution formulations by up to 75% when compared to a single orifice with an equivalent cross sectional area (p<0.05). While prior work has relied on metal actuator components, improvements in micro injection moulding and micro drilling now make it possible to mass produce novel orifice shapes to achieve similar FPF gains in plastic parts, with orifice diameters less than 0.2 mm. The ability to create internal features inside the actuator is also demonstrated. We show through in vitro high speed imaging that twin orifice sprays merge quickly and act as a single, modified plume. We also show for the first time that FPF and fine particle dose (FPD) are strongly correlated with the distance at which the plume velocity decays to half its initial value (R2=0.997 and 0.95 respectively). When plume velocity & FPF are increased, mouthpiece deposition decreases. This suggests that while smaller orifices produce more fine particles, higher sustained plume velocities also entrain more of the fine particles produced at the periphery of the spray due to increased shear. The effect occurs within the mouthpiece and is thus unlikely to alter the flow field in the upper airway.


Subject(s)
Metered Dose Inhalers , Nebulizers and Vaporizers , Administration, Inhalation , Aerosols , Equipment Design , Particle Size
4.
Int J Pharm ; 566: 463-475, 2019 Jul 20.
Article in English | MEDLINE | ID: mdl-31173800

ABSTRACT

This paper presents in situ time-resolved drug mass fraction measurements in pressurised metered dose inhaler (PMDI) sprays, using a novel combination of synchrotron X-ray fluorescence and scattering. Equivalent suspension and solution formulations of ipratropium bromide in HFA-134a propellant were considered. Measurements were made both inside the expansion chamber behind the nozzle orifice, and in the first few millimeters of the spray where droplet and particle formation occur. We observed a consistent spike in drug mass fraction at the beginning of the spray when the first fluid exits the nozzle orifice. Approximately 20% of the total delivered dose exits the nozzle in the first 0.1 s of the spray. The drug mass fraction in the droplets immediately upon exiting the nozzle peaked at approximately 50% of the canister mass fraction, asymptoting to approximately 20% of the canister concentration. The effect is due to a change in the drug mass fraction inside the droplets, rather than changes in droplet size or distribution. The transient was found to originate inside the expansion chamber. We propose that this effect may be a major contributor to low delivery efficiency in PMDIs, and have important implications for oropharyngeal deposition and inhalation technique. This highlights the importance of expansion chamber and nozzle design on the structure of PMDI sprays, and indicates areas of focus that may lead to improvement in drug delivery outcomes.


Subject(s)
Metered Dose Inhalers , Aerosol Propellants/chemistry , Bronchodilator Agents/chemistry , Equipment Design , Hydrocarbons, Fluorinated/chemistry , Ipratropium/chemistry , Pressure , Solutions , Spectrometry, X-Ray Emission , Suspensions
5.
Pharm Res ; 34(4): 718-729, 2017 04.
Article in English | MEDLINE | ID: mdl-28097509

ABSTRACT

PURPOSE: Sprays from pressurised metered-dose inhalers are produced by a transient discharge of a multiphase mixture. Small length and short time scales have made the investigation of the governing processes difficult. Consequently, a deep understanding of the physical processes that govern atomisation and drug particle formation has been elusive. METHODS: X-ray phase contrast imaging and quantitative radiography were used to reveal the internal flow structure and measure the time-variant nozzle exit mass density of 50 µL metered sprays of HFA134a, with and without ethanol cosolvent. Internal flow patterns were imaged at a magnification of 194 pixels/mm and 7759 frames per second with 150 ps temporal resolution. Spray projected mass was measured with temporal resolution of 1 ms and spatial resolution 6 µm × 5 µm. RESULTS: The flow upstream of the nozzle comprised large volumes of vapour at all times throughout the injection. The inclusion of ethanol prevented bubble coalescence, altering the internal flow structure and discharge. Radiography measurements confirmed that the nozzle exit area is dominantly occupied by vapour, with a peak liquid volume fraction of 13%. CONCLUSION: Vapour generation in pMDIs occurs upstream of the sump, and the dominant volume component in the nozzle exit orifice is vapour at all times in the injection. The flow in ethanol-containing pMDIs has a bubbly structure resulting in a comparatively stable discharge, whereas the binary structure of propellant-only flows results in unsteady discharge and the production of unrespirable liquid masses.


Subject(s)
Equipment Design/instrumentation , Ethanol/chemistry , Metered Dose Inhalers , Chemistry, Pharmaceutical , Humans , Light , Pressure , Radiography , X-Rays
6.
Pharm Res ; 33(5): 1249-58, 2016 May.
Article in English | MEDLINE | ID: mdl-26887680

ABSTRACT

PURPOSE: Typical methods to study pMDI sprays employ particle sizing or visible light diagnostics, which suffer in regions of high spray density. X-ray techniques can be applied to pharmaceutical sprays to obtain information unattainable by conventional particle sizing and light-based techniques. METHODS: We present a technique for obtaining quantitative measurements of spray density in pMDI sprays. A monochromatic focused X-ray beam was used to perform quantitative radiography measurements in the near-nozzle region and plume of HFA-propelled sprays. RESULTS: Measurements were obtained with a temporal resolution of 0.184 ms and spatial resolution of 5 µm. Steady flow conditions were reached after around 30 ms for the formulations examined with the spray device used. Spray evolution was affected by the inclusion of ethanol in the formulation and unaffected by the inclusion of 0.1% drug by weight. Estimation of the nozzle exit density showed that vapour is likely to dominate the flow leaving the inhaler nozzle during steady flow. CONCLUSIONS: Quantitative measurements in pMDI sprays allow the determination of nozzle exit conditions that are difficult to obtain experimentally by other means. Measurements of these nozzle exit conditions can improve understanding of the atomization mechanisms responsible for pMDI spray droplet and particle formation.


Subject(s)
Aerosol Propellants/chemistry , Bronchodilator Agents/administration & dosage , Hydrocarbons, Fluorinated/chemistry , Ipratropium/administration & dosage , Metered Dose Inhalers , Equipment Design , Volatilization , X-Rays
7.
Pharm Res ; 33(4): 816-25, 2016 Apr.
Article in English | MEDLINE | ID: mdl-26563204

ABSTRACT

PURPOSE: Drug concentration measurements in MDI sprays are typically performed using particle filtration or laser scattering. These techniques are ineffective in proximity to the nozzle, making it difficult to determine how factors such as nozzle design will affect the precipitation of co-solvent droplets in solution-based MDIs, and the final particle distribution. METHODS: In optical measurements, scattering from the constituents is difficult to separate. We present a novel technique to directly measure drug distribution. A focused x-ray beam was used to stimulate x-ray fluorescence from the bromine in a solution containing 85% HFA, 15% ethanol co-solvent, and 1 [Formula: see text] / [Formula: see text] IPBr. RESULTS: Instantaneous concentration measurements were obtained with 1 ms temporal resolution and 5 [Formula: see text] spatial resolution, providing information in a region that is inaccessible to many other diagnostics. The drug remains homogeneously mixed over time, but was found to be higher at the centerline than at the periphery. This may have implications for oropharyngeal deposition in vivo. CONCLUSIONS: Measurements in the dynamic, turbulent region of MDIs allow us to understand the physical links between formulation, inspiration, and geometry on final particle size and distribution. This will ultimately lead to a better understanding of how MDI design can be improved to enhance respirable fraction.


Subject(s)
Metered Dose Inhalers , Pharmaceutical Preparations/analysis , Spectrometry, X-Ray Emission/methods , Aerosol Propellants/chemistry , Bromine/chemistry , Equipment Design , Ethanol/chemistry , Hydrocarbons, Fluorinated/chemistry , Solvents/chemistry
8.
Pharm Res ; 31(11): 2963-74, 2014 Nov.
Article in English | MEDLINE | ID: mdl-24934661

ABSTRACT

PURPOSE: Non-volatile agents such as glycerol are being introduced into solution-based pMDI formulations in order to control mean precipitant droplet size. To assess their biopharmaceutical efficacy, both microscopic and macroscopic characteristics of the plume must be known, including the effects of external factors such as the flow generated by the patient's inhalation. We test the hypothesis that the macroscopic properties (e.g. spray geometry) of a pMDI spray can be predicted using a self-similarity model, avoiding the need for repeated testing. METHODS: Glycerol-containing and glycerol-free pMDI formulations with matched mass median aerodynamic diameters are investigated. High-speed schlieren imaging is used to extract time-resolved velocity, penetration and spreading angle measurements of the pMDI spray plume. The experimental data are used to validate the analytical model. RESULTS: The pMDI spray develops in a manner characteristic of a fully-developed steady turbulent jet, supporting the hypothesis. Equivalent glycerol-containing and non glycerol-containing formulations exhibit similar non-dimensional growth rates and follow a self-similar scaling behaviour over a range of physiologically relevant co-flow rates. CONCLUSIONS: Using the proposed model, the mean leading edge penetration, velocity and spreading rate of a pMDI spray may be estimated a priori for any co-flow conditions. The effects of different formulations are captured in two scaling constants. This allows formulators to predict the effects of variation between pMDIs without the need for repeated testing. Ultimately, this approach will allow pharmaceutical scientists to rapidly test a number of variables during pMDI development.


Subject(s)
Aerosols/chemistry , Pharmaceutical Solutions/chemistry , Technology, Pharmaceutical/methods , Administration, Inhalation , Chemistry, Pharmaceutical/methods , Metered Dose Inhalers , Particle Size